Keithley 2520 Service Manual page 44

Pulsed laser diode test system
Hide thumbs Also See for 2520:
Table of Contents

Advertisement

2-14
Calibration
Step 4: Pulse low calibration
Follow the steps below to calibrate current source pulse low.
calibration currents and test resistance values.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
Table 2-6
Pulse low calibration values
Current source range Calibration values
500mA
5A
Connect DMM and 100Ω resistor to the Model 2520 testhead CURRENT
OUTPUT jacks, as shown in
Select the DMM DC volts function, and enable auto-range.
Press the LASER I
key, press the EDIT key, then set the source to the 500mA
L
range using the RANGE key.
Press the MENU key, select CAL, then press ENTER.
Select EXECUTE, then press ENTER to enter the calibration menu.
Select PULS_LOW, then press ENTER. The unit displays:
Pulse Low CAL
Press ENTER to Output +15.000mA
Press ENTER. The unit displays:
DMM RDG: +15.000mA
Use ,
, , , ENTER or EXIT.
Note the DMM voltage reading, then calculate the current from the voltage and
actual resistance value: I = V/R.
Adjust the display to agree with the calculated current, then press ENTER. The
unit displays:
Pulse Low CAL
Press ENTER to Output +01.500mA
Press ENTER. The unit displays:
DMM RDG: +01.500mA
Use ,
, , , ENTER or EXIT.
Again, note the DMM voltage reading, then calculate the current from the voltage
and actual resistance value: I = V/R.
15mA, 1.5mA
150mA, 15mA
Model 2520 Pulsed Laser Diode Test System Service Manual
Figure
2-4.
Calibration resistance
100Ω
10Ω
Table 2-6
summarizes