Keithley 2520 Service Manual page 42

Pulsed laser diode test system
Hide thumbs Also See for 2520:
Table of Contents

Advertisement

2-12
Calibration
6.
7.
8.
9.
10.
11.
12.
Figure 2-2
Voltage measurement calibration connections
Model 2520 Testhead
CURRENT
BIAS
INPUT
DETECTOR 1
DETECTOR 2
!
ISOLATION FROM EARTH: 10V MAX.
VOLTAGE
SENSE HI
Step 3: Current source calibration
Follow the steps below to calibrate the current source.
resistors.
1.
2.
3.
4.
Connect the calibrator to the Model 2520 testhead VOLTAGE SENSE jacks, as
shown in
Figure
2-2. (Connect calibrator HI to VOLTAGE SENSE LO; connect
calibrator LO to VOLTAGE SENSE HI.)
Turn on the calibrator output, set the calibrator voltage to -10.000V, then press
ENTER. The unit will prompt you as follows:
VL-CAL
Switch HI and LO connections
Switch the HI and LO calibrator connections, then press ENTER.
Press EXIT to return to normal display.
Press the LASER V
key to make sure the laser volts mode is selected.
L
Press the RANGE key to select the 5V range.
Repeat steps 2 through 9 for the 5V range. Be sure to set the calibrator output to
-5.000V as shown in
Table
HI
LO
CURRENT
OUTPUT
!
CAT I
VOLTAGE
SENSE
HI
LO
BNC Cables
VOLTAGE
SENSE LO
From normal display, press the Model 2520 LASER V
10V range using the RANGE key.
Press the LASER I
key, press the EDIT key, then set the source to the 500mA
L
range using the RANGE key.
Press the MENU key, select CAL, then press ENTER.
Select EXECUTE, then press ENTER to enter the calibration menu.
Model 2520 Pulsed Laser Diode Test System Service Manual
2-4.
Calibrator (Output DC Voltage)
Output
HI
Output LO
Table 2-5
summarizes calibration
key, then set the range to
L