NOTE:
The supported USB color sensor for SENS/MEM port is MDSVSENSOR3.
After the monitor is powered on, the "Color stabilizer" function is busy internally and must be warmed up. Calibrating
during this period will affect the calibration quality. For the best calibration results, wait 30 minutes for the monitor to
warm up before starting the calibration.
The results of the "Self calibration" and "Validation" functions are stored in the monitor and can be read by software
on your computer. Setting the monitor's time clock is required for this function. Follow the instructions on the OSD
menu and set the time. After the time is set, the monitor counts automatically while the AC power is active.
"White copy" will adjust only luminance and white point. For more accurate color matching please use the MultiProfiler
software, which can be download from the NEC Display Solutions website.
English-26