Adjustment Report Printout; Determining Printout Content; Adjustment Report; Header, Footer, Glp Printouts - RADWAG AS R2 User Manual

R series balances
Hide thumbs Also See for AS R2:
Table of Contents

Advertisement

7.4. ADJUSTMENT REPORT PRINTOUT

At the end of each adjustment process or adjustment test, an adjustment report is generated
automatically and sent to a communication port COM 1. The content of the report is declared in menu
P5.1 CAL REPORT.
The instruction on how to declare settings for this option is described in section about printouts.
The report can be printed out via a printer connected to the balance or it can be sent to the computer
and saved as a file for archiving purposes.

8. DETERMINING PRINTOUT CONTENT

8.1. ADJUSTMENT REPORT

P5.1 CAL. REPORT, is a group of parameters enabling user to declare data that is to be printed on
an adjustment printout.
Variable
PROJECT
CALIB TYPE
USER
PROJECT
DATE
TIME
BALANCE ID
CAL. DIFFER
DASHES
SIGNATURE
For the parameters described above, one of these values must
be selected:
NO - do not print
YES - print

8.2. HEADER, FOOTER, GLP PRINTOUTS

HEADER
GLP PRINTOUT
Overview
Option enables naming the project (name associated with a particular type of
weighing). The name may consist of 16 characters maximum.
Option enables printing out the type of the adjustment being carried out.
Option enables printing out the name of a logged-in user.
Option enables printing out the name of the project (see parameter Project).
Option enables printing out the date of the carried out adjustment.
Option enables printing out the time of the carried out adjustment.
Option enables printing out the balance ID number.
Option enables printing out the difference between mass of an adjustment
weight measured during the last adjustment and the current measured mass
of this weight.
Option enables printing out dashes that separate the date of a printout from
a signature.
Option enables providing an area for the signature of a user performing the
adjustment.
group of parameters enabling to declare data that is to be printed on
a header printout.
group of parameters enabling to declare data that is to be printed on
a measurement result printout
An exemplary report: