V2 (two-point voltage
calibration)
22
Calibration
Point 1 (P1) - The DPI 104 shows the following displays:
1. The calibration point to be used for V2 - Point 1.
8 seconds
After step 2, the DPI 104 sets Vout to 0.1 V. Correct the value
(P1) to the value shown on the voltage calibrator.
5. Repeat steps 3. & 4. for each digit. The value is ignored if it is
not in the permitted limits (Table 11, page 19).
Point 2 (P2) - If P1 has a permitted value, the DPI 104 shows these
displays:
1. The calibration point to be used for V2 - Point 2.
8 seconds
After step 2, the DPI 104 sets Vout to 5.0 V. Correct the value
(P2) to the value shown on the voltage calibrator.
5. Repeat steps 3. & 4. for each digit. The value is ignored if it is
not in the permitted limits (Table 11, page 19).
6. If P2 has a permitted value, the DPI 104 uses the new P1/P2
values to adjust the output (Vt):
• Press this button to accept the V2 calibration.
• The display shows "donE".
• The instrument will restart.
• Press this button to reject the V2 calibration and move to the next
menu option.
2.
3.
0 to 9
2.
3.
0 to 9
4.
4.
K0394 Issue 3 - [EN] English