Menu: Calibration

Menu:
Set low/high
alarm
Low/high alarm - Set-up
and use
12
Operation
The switch symbol and the value on the main display flash to give
the switch condition and the switch pressure.
Press
To reset the monitor function.
Refer to the menu operations section.
Use the alarm function to show when the pressure is not in the
specified limits for the system.
Set applicable values in the range 0 to 105% FSO:
%FSO = [ Applied Pressure / (FSO High — FSO Low) ] x 100
Note: When setting a tare value, the alarm function uses the
calibrated range and the pressure value on the display (refer to menu
operation, page 5).
The alarm indication is available on the display and as a signal
output (Table 6). Figure B2 (b) shows an example configuration with
Table 3 accessories B and C, page 4.
While there is an alarm condition, the applicable alarm symbol
(high or low) flashes on the display (A2: item 15).
Recommendation
To save battery power use an external power supply.
Menu: Set function to On (refer to menu operation, page 5). Then
use these steps to set the low and/or high alarm.
Alarm
(% FSO)
Digit = 0 or 1
5. To finish, repeat steps 3 & 4 for each digit.
If the value entered is not correct, the value resets to the nearest
permitted value. That is:
a value in the range 0 to 105% FSO
a low alarm value < high alarm value
To accept or change the new value, repeat steps 1 to 5.
Press
To cancel the new value.
Alarm
1
2
(% FSO)
Digit = 0 to 9
3
4
K0394 Issue 3 - [EN] English