Lock code - Set-up and
use
Menu: Set scan rate
Scan rate - Set-up and
use
Menu:
Monitor
external IDOS
Monitor external IDOS -
Set-up and use
[EN] English - K0394 Issue 3
Menu: Set to On (refer to menu operation, page 5). Use the steps
that follow to set a new code.
Lock
Digit = 0 or 9
3. Complete the lock code, repeat steps 1 & 2 for each digit.
4. To change the menu options, the display shows: L - - -
5. Enter the applicable code.
To reset the code to the factory code, do a restore operation. Refer
to menu operations, page 5.
This function sets the rate that the DPI 104 uses to take pressure
samples from the internal sensor.
The nominal update rate for the display is 2 Hz. The up-date rate for
the maximum/minimum function and the switch function is greater
than or equal to the specified scan rate.
Note: Increasing the scan rate, increases the power consumption.
Menu: Set an applicable value in the range 2 to 10 Hz, refer to menu
operation, page 5.
Use this function to read the pressure from an external IDOS UPM.
All the other DPI 104 pressure functions are available but not the
calibration function.
Example: Set tare (or set zero), Monitor maximum/minimum.
This function does not supply power to the IDOS UPM and needs
accessories B and C, refer to Table 3, page 4.
1. Connect the DPI 104 (Figure B2 (d)).
2. Menu: Set to On (refer to menu operation, page 5).
1
2
Lock
Digit = 0 to 9
15
Operation