[EN] English - K0394 Issue 3
External software can be used with the DPI 104 (Table 3
accessories A or D, page 4). If SiCal PRO (accessory D) is used, all the
menu commands and display data are available.
Note: To use SiCal PRO, it must have the serial number of each unit
to be used with (DPI 104, IDOS UPM, PC6-IDOS).
This symbol appears when the DPI 104 transmits or
receives data.
Set-up a network of up to 99 units in series ('daisy chain'). Figure 4
shows the electrical connections to do this, refer to Table 6 page 8.
12-24 Vdc
(+)
GND
DPI 104 (1)
Figure 4: Connections for a DPI 104 network
Table 9: Error codes/indications
Code
Description
E0001
Incorrect unlock code.
E0002
The tare facility is not available
because the menu lock is On and
the lock code < 500.
E0004
Start-up error.
E0005
External IDOS UPM not found.
E0006
Incorrect calibration access code.
E0007
The power supply voltage is too low
to do a calibration.
E0009
Unable to supply the specified Vout. Example:
Applied pressure ≥ 110% FS.
OLoAd
99999/
There are not enough digits in the
-9999
main display to give the correct
pressure value.
PC: 9-way, D-type
connector
GND
Tx
Rx
DPI 104 (2)
Action
Use the correct code.
Change the menu configuration.
Do a restore operation (refer to
menu operation, page 5).
Make sure that all the related
equipment and connections are
serviceable.
Use the correct code.
Use an external power supply or
replace the battery.
Low battery. Use an external power
supply or replace the battery.
Bad connection. Make sure that all
the related equipment and
connections are serviceable.
Reduce the pressure.
Change the measurement units.
17