GE druck dpi 104 User Manual page 12

Digital pressure indicator
Hide thumbs Also See for druck dpi 104:
Table of Contents

Advertisement

Tare - Set-up and use
Tare - With Lock
Tare - With Alarm
and/or Vout
Tare - With FSO values
10
Operation
Menu: Set this function to On (refer to menu operation, page 5).
When this function is On, there are two options to set a
tare value (tA):
Menu option: Set the menu "t On", then set a tA value:
A display
t
0 to 9, or -
Repeat steps 1 & 2 for each digit and for the decimal point.
Zero option: Step 1 lets you set a value for tA. Press and hold.
Normal
output
A = 0
t
When tA is not zero, the last segment on the analog dial flashes.
To make sure that there is an indication of the correct pressure:
While tare is On, the analog dial and % indication show values
calculated from the calibrated range without the tare adjustment.
If the menu lock is On with a lock code set < 500, the zero option is
rejected - Error code (E0002).
If the zero option is used to set a tare value (tA) while the alarm
and/or Vout functions are On, the display counts down from:
tArE9 to tArE0.
To cancel the specified tA value, press this button.
To continue with the specified tA value, press this button OR
let the count complete.
When setting a tA value, the alarm and Vout functions use values
calculated from the calibrated range and the pressure value on the
display.
To make sure that there is an indication of the correct pressure
while tare is On, the FSO Low and/or FSO High values are not used.
1
2
A display
t
0 to 9
Normal
1
output
donE
A = 70 mbar
t
K0394 Issue 3 - [EN] English

Hide quick links:

Advertisement

Table of Contents
loading