Performing Calibration Correction (/Cc1 Option) - YOKOGAWA Daqstation DX2000 User Manual

Hide thumbs Also See for Daqstation DX2000:
Table of Contents

Advertisement

3.9
Performing Calibration Correction (/CC1
Option)
Setup Screen
Setup Items
3-16
The input value is corrected using segments, and the result is used as a measured value.
For a description of the function, see section 1.1.
Press MENU (to switch to setting mode), and select the Menu tab > Meas channel >
Calibration correction.
Y5
Y4
MES val
Y3
Y2
Y1
X1 X2
X3 X4
• First-CH/Last-CH
Select the target channels. You can set consecutive channels whose range is set to
the same value as the first channel.
• Calibration correction > Number of set points
Select the number of points that make up the segments (including the start and end
points) in the range of 2 to 16.
To disable calibration correction, select Off.
• Calibration correction > MES val, True val
Press the Input soft key and enter the value.
For the MES value, set a value that is greater than the previous value.
Press the Measure soft key to set the measured value at that point to MES val. If
you press the Measure soft key when setting multiple channels simultaneously, the
measured value of the first channel are set to the MES val of all channels.
Selectable Range of MES and True Values
• Channels on which linear scaling is specified
–30000 to 30000 (the decimal place is the same setting as the scale value)
• Other channels
Value in the measurable range of the selected range
Example: –2.0000 to 2.0000 for 2 V range
Note
• The calibration correction setting is set to Off if you change the Mode or Range setting.
• Calibration correction cannot be specified on channels set to Skip.
Condition
X1 < X2 < X3 < X4 < X5
Calibration set points
Number of set points
True val
X5
IM 04L42B01-01E