Cerabar T
Power
Switch frequency
Input PLC
Inductive loads
PMC131
Endress+Hauser
max. 6 W
max. 10 Hz
≤ 2 kΩ
• Input resistance R
i
≥ 10 mA
• Input current I
i
To prevent electrical interference, only operate an inductive load (relays, contactors, solenoid valves) when
directly connected to a protective circuit (free-wheeling diode or capacitor).
Power supply
Plug/cable connection
1
2
ENDRESS+HAUSER
CERABAR T
M16 / ½ NPT
Plug M 16 x 1.5 (DIN 43650/A), ½ NPT Plug M 12x1
➀ Plug-in housing
➁ Gasket
Electrical connection: Analog/current output
1
4...20 mA
1
2
2–
1+
1
Plug M 16 x 1.5 (DIN 43650/A), ½ NPT
2
Plug M 12 x 1
3
Cable (rd = red, bk = black, gnye = green-yellow)
ENDRESS+HAUSER
CERABAR T
M12x1
2
+
12...30 V
–
3-
1+
3
PE
1
ENDRESS+HAUSER
CERABAR T
5 m/25 m
P01-PMC131xx-04-xx-xx-xx-001
5 m/25 m cable
➀ Reference pressure line
+
12...30 V
–
4...20 mA
2
+
12...30 V
–
4...20 mA
P01-PMC131xx-04-xx-xx-xx-002
5