Pmp131; Supply Voltage; Residual Ripple; Cable Entry - Endress+Hauser Cerabar T PMC131 Technical Information

Pressure transducer, with ceramic and metal sensors for absolute pressure and gauge pressure measurement up to 400 bar
Hide thumbs Also See for Cerabar T PMC131:
Table of Contents

Advertisement

Cerabar T

PMP131

Supply voltage

Residual ripple

Cable entry

Endress+Hauser
Electrical connection (switch output)
R
1
1
1+
2
3
1
Plug M 16 x 1.5 (DIN 43650/A), ½ NPT
2
Plug M 12 x 1
R
External load, e.g. relay, programmable logic controller, distributed control system
For electrical connection provided by customer use only shielded cable
Electrical connection: Analog-/voltage output
0...10 V
1
1+
2-
3+
1
Plug M 16 x 1,5 (DIN 43650/A), ½ NPT and plug DIN 43 650/C
2
Plug M 12 x 1
3
Cable (rd = red, wh = white, gn = green)
For electrical connection provided by customer use only shielded cable
PMC131
11 to 30 V DC
PMP131 and PMP135 (current output, 2-wire version)
• For non-hazardous areas: 12 to 30 V DC
• Ex i: no-load voltage ≤ 26 V DC, short-circuit current ≤ 100 mA, power consumption ≤ 0.8 W
PMP131 (voltage output, 3-wire version)
• 15...30 V DC
PMP131 and PMP135 (switch output)
• 18 to 32 V DC, current consumption without load < 20 mA, with reverse polarity protection
• Analog output: max. 5 % of supply voltage
• Switch output: max. 10 % of supply voltage
→ See Page 18, "Ordering information" section.
+
18...32 V
3+
2 –
2
+
15...30 V
3-
2+
1+
2
+
18...32 V
R
0...10 V
3
+
15...30 V
PE
3
2-
1+
P01-PMP13xxx-04-xx-xx-xx-003
0...10 V
+
15...30 V
P01-PMP131xx-04-xx-xx-xx-001
7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cerabar t pmp131Cerabar t pmp135