Endress+Hauser Cerabar T PMC131 Technical Information page 16

Pressure transducer, with ceramic and metal sensors for absolute pressure and gauge pressure measurement up to 400 bar
Hide thumbs Also See for Cerabar T PMC131:
Table of Contents

Advertisement

16
Switch output: Switch point and hysteresis adjustment
1
+
2
1
Hysteresis adjustment 1.5 to 20 % of URL;
Factory setting 10 % of URL
2
Switch point adjustment 0 to 100 % of URL;
Factory setting 50 % of URL
3
LED color code for checking the switch status:
green = off; red = on
4
Connect voltmeter to test pins:
0 to 1 V corresponds to 0 to 100 % of URL
U
out
ON = red LED
on
off
Instructions for switch point and hysteresis (the percentage values refer to the URL)
U
Output voltage
a
p
Acting pressure
For the Cerabar T PMP131 and PMP135 with
switch output, both the switch point and hysteresis
can be adjusted.
This can also be carried out at atmospheric pressure
3
using the enclosed test cable and a voltmeter.
4
P01-PMx13xxx-19-xx-xx-xx-002
Hysteresis
1.5...20 %
Switch
on
Switch point
0...100 %
Cerabar T
Switch
off
OFF = green LED
p
P01-PMP13xxx-05-xx-xx-en-001
Endress+Hauser

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cerabar t pmp131Cerabar t pmp135

Table of Contents