ResMed C16389 User Manual page 44

Resmed mask specification sheet
Table of Contents

Advertisement

Available languages

Available languages

268106r1.book Page 39 Thursday, November 30, 2006 2:35 PM
Guidance and manufacturer's declaration – electromagnetic immunity (Continued)
The VPAP Malibu is intended for use in the electromagnetic environment specified below. The customer or
the user of the VPAP Malibu should assure that it is used in such an environment.
Immunity
IEC60601-1-2
test
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to 2.5
GHz
NOTE 1: At 80 MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the VPAP Malibu is used exceeds the applicable RF compliance level above, the VPAP Malibu should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the VPAP Malibu.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Compliance
Electromagnetic environment –
level
guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the VPAP Malibu, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
3 Vrms
d = 1.17 P
10 V/m
d = 0.35 P 80 MHz to 800 MHz
d = 0.70 P 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
a
survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
b
Technical Specifications
39