INTEGRATED HUMIDIFIER SETTING UP THE S8 ™ ™ S8 AutoSet Spirit System ILLUSTRATIONS Component of 338193/1 SYSTEM COMPONENTS HumidAire 3i ™ Carry handle Keypad screen OTHER ACCESSORIES Air outlet Air filter cover Module Air inlet DC input AC input Note: Some products not available in all regions...
Page 2
HUMIDAIRE 3i AND S8 USING THE DATA CARD HumidAire 3i BLANK MODULE Data communications port DATA CARD MODULE Note: Some products not available in all regions...
NTRODUCTION Thank you for choosing the flow generator. The ™ S8 A S8 A PIRIT is a compact, all-in-one package with an inbuilt power supply and the option of PIRIT integrated data management and humidification. This user manual contains the information you need for the correct use of your S8 A PIRIT ESPONSIBILITY...
A patient should not connect a device to the data communication port unless • instructed to do so by their health care provider or physician. Only ResMed products are designed to be connected to the data communication port. Connecting other devices could result in injury, or damage to the S8 A device.
Caution should be exercised if the room temperature is warmer than 90 F (32 When AC mains power (100/240 V AC) is not available, always use a ResMed • DC-12 converter. (The DC-12 converter is available as an optional accessory.)
The 2 m air tubing connects the to your mask. S8 A PIRIT ASKS You will also need a ResMed mask system (supplied separately). The ResMed mask systems that are recommended for use with the S8 A PIRIT are: ASAL ASKS •...
IRAGE Note: Not all masks are available in all regions. WARNING In AutoSet mode, only ResMed mask systems are compatible for use with the S8 AutoSet Spirit. Please refer to “Warnings” on page 2. UMIDIFIER Please refer to the illustration in section B of the illustration sheet.
Note: You can also place your S8 AutoSet Spirit on the floor beside or underneath your bed. Ensure that the area is free from dust and clear of bedding, clothes or any other objects that could block the air inlet.
TART REATMENT Make sure the power is on. A welcome message is displayed briefly on the LCD screen, then the standby screen appears. In CPAP mode, the standby screen displays the word “RAMP”. In AutoSet mode, the standby screen displays the word “SETTLING”. The key and LCD backlights also turn on.
TOPPING REATMENT To stop treatment at any time, remove your mask and press the Start/Stop key if your clinician has enabled SmartStart, simply remove your mask and treatment will end. SING THE ITTING EATURE You can use the S8 mask-fitting feature to help you fit your mask PIRIT properly.
Page 16
After three minutes, the pressure reverts to the set pressure and treatment will begin. If you do not wish to wait three minutes, hold down the Start/Stop key for at • least three seconds to start treatment immediately. If you press the Start/Stop key for less than three seconds, the unit will return •...
Make sure the water chamber of the HumidAire 3i is empty before you attach or detach the humidifier. b. Remove the connector plug on the S8 AutoSet Spirit before you attach the HumidAire 3i. Reattach the connector plug firmly after you detach the humidifier.
• Starts or stops treatment. • Extended hold (for at least three seconds) starts the mask- fitting feature. • Allows you to scroll through the S8 AutoSet Spirit menus and submenus, and increase the setting options. Down • Allows you to scroll through the S8 AutoSet Spirit menus and submenus, and decrease the setting options.
Page 20
When the device is in standby mode all keys remain lit. During therapy, the keypad backlight turns off after two minutes of inactivity and the Start/Stop key dims. OW TO HANGE THE ETTINGS You can change settings on the by using the keypad and LCD. S8 A PIRIT When the...
Page 21
Selects the ramp time. OFF–45 min; 5-minute (CPAP Mode Ramp time is the period during increments only) which the S8 AutoSet Spirit increases from a low pressure to the prescribed treatment pressure. Settling Selects the settling time. OFF–45 min; 5-minute...
OW TO SE THE TANDARD Figure 1 shows the standard menu. CPAP AUTOSET WELCOME WELCOME >>>>>> >>>>>> RAMP: 10min SETTLING: 10min menu menu EPR SETTING: 1 change exit exit USAGE: 1000/1068 exit exit USAGE: 1000/1068 MASK: ULTRA change exit exit MASK: ULTRA SW: SX3960200 exit...
Page 24
The following table shows the setting that should be selected for each mask type. Table 3: Settings for mask types Settings Mask ULTRA Ultra Mirage Nasal Mask MIR FULL Mirage Full Face Mask Mirage Full Face Mask Series 2 Ultra Mirage Full Face Mask ACTIVA Mirage Activa Nasal Mask SWIFT...
OW TO SE THE ETAILED Figure 2 shows the detailed menu. CPAP AUTOSET WELCOME WELCOME >>>>>> >>>>>> SETTLING: 10min RAMP: 10min menu menu CPAP AUTOSET SETTINGS enter exit MASK: ULTRA change exit change exit MASK: ULTRA TUBE LENGTH: 2m change exit change exit...
Page 26
OW TO HANGE THE XPIRATORY RESSURE ELIEF ETTING Note: EPR may not be available in all regions. In CPAP mode, your clinician may have set up your device to deliver expiratory pressure relief (EPR). This can be used to help ease any discomfort you may experience when breathing out during treatment.
to change the option to ON or OFF. OW TO HANGE THE LERT ETTING Leak alert indicates when a high mask leak is present. Scroll to LEAK ALERT and select Change (Left key). Press the Up or Down keys to change the option to ON or OFF.
Page 28
Your clinician can set any of the following reminders on your LCD: Message Description Action INSERT CARD May appear if your Insert your Data Card (see Section H of the device is Data Card illustrations sheet) and follow any instructions enabled.
S8 M ODULES Note: Do not remove modules while power is connected to the S8 AutoSet Spirit. Your will have one of two modules fitted to the back: a blank S8 A PIRIT module or a Data Card module. SING A...
Page 30
NSERT THE Switch on the and wait until you see the standby (RAMP or S8 A • PIRIT SETTLING) screen. Hold the Data Card with the arrow facing up and insert it into the slot in the • Data Card module (H-1). Push the card in smoothly until it stops moving (H-2) and messages start •...
Page 31
WARNING If your clinician has told you to use this Data Card to update the settings on your device and the “Settings Success” message does not appear, contact your clinician immediately. S8 M ODULES...
LEANING AND AINTENANCE You should regularly carry out the cleaning and maintenance described in this section. Refer to your mask manual for detailed instructions. AILY LEANING Clean the mask according to the instructions supplied with the mask. IR TUBING Disconnect the air tubing from the (and humidifier, if used) and hang S8 A PIRIT...
Ensure the air filter and air filter cover are fitted at all times. ERVICING CAUTION Do not attempt to open the S8 AutoSet Spirit case. There are no user • serviceable parts inside. Inspection and repair should only be performed by an authorised agent. Under •...
REQUENTLY SKED UESTIONS S8 A OW WILL IT FEEL WHEN AM BREATHING WITH THE PIRIT When you first use your , you may feel uncomfortable breathing S8 A PIRIT against the airflow. This is normal. You will continue to breathe normally while you sleep.
Page 36
You must use a ResMed DC-12 converter to connect the to a 12 V S8 A PIRIT or 24 V DC power source. Contact your equipment supplier or ResMed for details. WARNING The device should not be connected to both AC and DC power sources simultaneously.
ROUBLESHOOTING If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device. Problem Possible Cause Solution No display. Power not connected. Ensure the appropriate power...
Page 38
Replace them. Air tubing not connected properly. Connect firmly at both ends. Air tubing kinked or punctured. Straighten or replace. S8 AutoSet Spirit unit SmartStart/Stop is disabled. Enable SmartStart/Stop. does not stop when you remove your mask. SmartStart is enabled...
Page 39
If this does not clear the message, disconnect the power cord and then reconnect it to restart the device. Display error message: Component failure. Return your S8 AutoSet Spirit for servicing. SYSTEM ERROR Call service! Display error message: You have experienced...
Page 40
You may have removed the Data Reinsert the Data Card and wait Card before settings were copied for the Settings Success to the S8 AutoSet Spirit. Remove Card message to appear on the LCD. The card is not a ResScan Data Only use a ResScan Data Card.
UMID 400 Hz; 2.5A <140 VA (110 W) (maximum power consumption). Instantaneous peak power consumption <340 VA. Refer to the DC-12 converter instructions for DC ratings. Only use the ResMed • DC-12 converter for DC input. Actual power consumption will vary, depending on factors such as the use of accessories, height above sea level and ambient temperature.
ODULE Dimensions (H x W x D): 49.7 mm x 42.2 mm x 12 mm EVICES WHERE APPLICABLE Housing Construction: Flame retardant engineering thermoplastic Environmental Conditions Operating Temperature: +5°C to +40°C Operating Humidity: 10%–95% non-condensing Storage and Transport Temperature: -20°C to +60°C Storage and Transport Humidity: 10%–95% non-condensing Operating Altitude: sea level to 2591 m Electromagnetic Compatibility...
Page 43
’ UIDANCE AND ANUFACTURER ECLARATION LECTROMAGNETIC MISSIONS AND MMUNITY Guidance and manufacturer’s declaration—electromagnetic emissions The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment. Electromagnetic environment—...
Page 44
Guidance and manufacturer’s declaration—electromagnetic immunity The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment. IEC60601-1-2 Electromagnetic Immunity test...
Page 45
Guidance and manufacturer’s declaration—electromagnetic immunity The S8 series of devices is intended for use in the electromagnetic environment specified below. The customer or the user of the S8 device should assure that the device is used in such an environment. IEC60601-1-2 Compliance Immunity test...
Page 46
Recommended separation distances between portable and mobile RF communications equipment and the S8 series of devices The S8 series of devices is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the S8 device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the S8 device as recommended below, according to the maximum output power of the communications equipment.
Excludes single-use devices. Note: Some models are not available in all regions. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does...