Quick Reference Guide
HYDRIM
®
L110w
Successful Instrument Washing
• Remove composite, cement and amalgam at chairside
• Do not put burs in Hydrim
• Do not allow instruments to dry for longer than one hour before
processing in Hydrim
• Load cassettes and baskets according to manufacturer's guidelines.
Do not overload.
• For cassettes and hinged instruments, use the heavy-duty cycle
• Place small / light items in the basket with lid
• Do not put anodized aluminum instruments in Hydrim
How to Use Hydrim
1. Turn the machine ON
2. Touch the screen to begin
3. Touch the unlock symbol
4. Load instruments / cassettes into the machine
5. Close the door
6. Touch the desired cycle. The door will be locked during the cycle.
7. When the "cycle complete" message appears, touch the unlock
symbol and unload.

Advertisement

loading

Summary of Contents for SciCan HYDRIM L110w

  • Page 1 Quick Reference Guide HYDRIM ® L110w Successful Instrument Washing • Remove composite, cement and amalgam at chairside • Do not put burs in Hydrim • Do not allow instruments to dry for longer than one hour before processing in Hydrim •...
  • Page 2: Cleaning The Filter

    Replacing HIP Cleaning Solution When the “no detergent” message appears, the HIP cleaning solution box needs to be changed (part-number CS-HIPL). Cleaning the Filter If there is debris on the filter, remove it, rinse under a tap and replace. Ensure that the filter is firmly in position.
  • Page 3: Error Messages

    Error Messages Error Cause Solution The water does not reach Call for service. CF1 Water the required temperature in Heating failure the specified time The water does not fill CF2 Chamber Check that the water the chamber within the Filling Failure hoses are not kinked.