Hydrim L110w Service Guide 95-109977 Rev 1.0
Copyright 2007 SciCan Ltd . All rights reserved.

Advertisement

Table of Contents
loading

Summary of Contents for SciCan HYDRIM L110w

  • Page 1 HYDRIM L110w ® • Service Manual Hydrim L110w Service Guide 95-109977 Rev 1.0 Copyright 2007 SciCan Ltd . All rights reserved.
  • Page 2: Table Of Contents

    Local: 416.445.1600 Toll Free: 1.800.667.7733 Fax: 416.445.2727 Hydrim is registered trademark of SciCan Ltd. HIP is trademark of SciCan Ltd. All other trademarks referred to in this manual are the property of their respective owners. Hydrim L110w Service Guide 95-109977-1…...
  • Page 3: Cycle Description

    Maximum relative humidity: 80% for temp up to 31°C / 88°F 50% for temp up to 40°C / 104°F Operating temperature range: -5°C to 40°C / 23-104°F Max. altitude: 2000m / 6561.7 ft. Mains supply: + / -10% of nominal Hydrim L110w Service Guide 95-109977-1…...
  • Page 4: Safety Information

    Hydrim L110w by a third party or for the use of equipment or parts manufactured by a third party, including lost profits, any commercial loss, economic loss, or loss arising from personal injury.
  • Page 5 Reporting It is vital for SciCan to learn of any problems in the field. This information will help SciCan solve the problem quickly and improve product reliability in new units. Biological Waste Wastewater in the unit may contain biological contaminants; use a mechanical means to siphon the contents.
  • Page 6 The maximum length of the drain hose is 3.3 m/13’. Installation Instructions The Hydrim L110w should only be installed and service by a qualified SciCan contractor, as it is an installation Category 2 device. The contractor should be experienced in installing equipment that requires electrical hook-up as well as plumbing.
  • Page 7 The unit must be connected to the water supply in accordance with all local and national plumbing codes. SciCan recommends a hard plumbing installation within 1.5m/5ft. of the unit. If additional distance is necessary, commercial grade plumbing hose must be used to minimize leaks.
  • Page 8 Once you have determined the water hardness level in ppm (parts per million) find the recommended setting for your Hydrim L110w in the water hardness setup table. If your water hardness falls between two settings, select the higher setting (see page 17 & 18 for changing the setting). Water softening salt (available in most supermarkets and hardware stores) should be added regularly.
  • Page 9: Routine Maintenance

    To add water softening salts follow these steps: Unscrew the salt container lid. Upon first use of the Hydrim L110w, pour approximately 1 liter (1 quart) of water into the salt container, or until it is full with water. It is not necessary to add water during subsequent refills of the salt container.
  • Page 10 Disconnect the hot and cold water inlet hoses. Make sure the hoses are clean and free of debris. Make sure the inlet valves are free of debris. Valves Dosing Pump Tubing Replace internal tubing on Dosing Pump. Remove cover and replace hose Hydrim L110w Service Guide 95-109977-1…...
  • Page 11: Operating Instructions

    The stage within the cycle is shown beneath the bar (ex. Draining, Filling, Rinsing). 24 min The estimated time remaining (in minutes) is always displayed. Please note that the time will increase or decrease depending on the projected cycle completion. Press stop to abort cycle. Hydrim L110w Service Guide 95-109977-1…...
  • Page 12 CF 2 Chamber Filling __________________________________________________________________________________ Cycle complete screen P2 Wash When the cycle is complete. The program name and “Cycle Complete” is displayed. Cycle Complete Press the OK button to return to the main menu screen. Hydrim L110w Service Guide 95-109977-1…...
  • Page 13: Overview Of Menus

    SH11MR311 User – Page 15 Technician - Page 21 Setup – Page 17 Set Drying Time Unit No. Error History Language Selection Cycle Count Set Regeneration Set Button Beep Screen Saver Temperature C/F LCD Contrast Hydrim L110w Service Guide 95-109977-1…...
  • Page 14: User Menu

    Choose from 1 Default 10 min to 20 min. Touch default if you wish to return to the factory default. Use the back arrow to return to the previous screen. Hydrim L110w Service Guide 95-109977-1…...
  • Page 15 Hydrim. Touch cycle count and the number of Error History cycles (completed and aborted) will be displayed. This cannot be reset. Touch the Cycle Count back arrow to return to the previous menu. Hydrim L110w Service Guide 95-109977-1…...
  • Page 16: Setup Menu

    If you have more than one Hydrim in your office, you can designate them Unit #1, Unit #2, etc. Use the up and down arrows to select the unit number. The default is 0. Touch the back arrow to return to the previous menu. Default 0 Hydrim L110w Service Guide 95-109977-1…...
  • Page 17 The user can select between having the unit beep or not beep whenever a screen button is touched. Turn the beep ON or OFF by using the up and down arrows. The default setting is ON. Touch the back arrow to return to the previous menu. Default: On Hydrim L110w Service Guide 95-109977-1…...
  • Page 18 The user can change the contrast on the display. Touch the up and down arrows to select more or less contrast on the display. The default is Touch the back arrow to return to the previous menu. Default Hydrim L110w Service Guide 95-109977-1…...
  • Page 19: Technical Service Menu Overview

    Salt Regeneration Valve – ON/OFF Dosing Pump – ON/OFF Dryer Motor + Heater – ON/OFF Hot Water Valve – ON/OFF Cold Water Valve – ON/OFF Rinse Aid Valve – ON/OFF (not used) Drain Pump – ON/OFF Hydrim L110w Service Guide 95-109977-1…...
  • Page 20 9 min ↑↓ (choose from 5 to 15 min) Set Rinse Time: P2 Wash 1 min ↑↓ (choose from 1 to 10 min) P3 Heavy Duty Wash 1 min ↑↓ (choose from 1 to 10 min) Hydrim L110w Service Guide 95-109977-1…...
  • Page 21 60ºC ↑↓ (choose from 30 to 93ºC) Repeater Mode: In the ON position, the unit will continuously run the selected cycle. Factory Default: Resets all values to factory default. Software Upgrade: To upgrade software via the SciCan Datalogger. Hydrim L110w Service Guide 95-109977-1…...
  • Page 22: Technical Service Menu

    Up arrow pressed, change to previous Technician menu screen. Diagnostic Tools Factory Default Down arrow pressed, change to next Technician menu screen. Cycle Settings SW Upgrade Press the back arrow to return to the previous menu. Hydrim L110w Service Guide 95-109977-1…...
  • Page 23 Press the down arrow to move to the next page / error. Press the X in the lower left hand corner to clear the error history. There must be 4 errors stored to clear the error history and the last error will remain. Hydrim L110w Service Guide 95-109977-1…...
  • Page 24 Up arrow button pressed, increase Diagnostic Tools Set Debug Screen value. Set debug screen Down arrow button pressed, decrease value. Default button pressed, set to default value. Default: Off Press the back arrow to return to the previous menu. Hydrim L110w Service Guide 95-109977-1…...
  • Page 25 Dryer heater is OFF Drain Pump=ON Dosing=OFF Drain pump is ON Salt V=OFF Dosing is OFF Salt valve is OFF The Debug cycle should be used when troubleshooting to view the IO status of components. Hydrim L110w Service Guide 95-109977-1…...
  • Page 26 Use the up and down arrows to select the desired drying time. Choose from 1 to 20 min. Press default Default 10 min if you wish to return to the factory default. Press the back arrow to return to the Technician Menu. Hydrim L110w Service Guide 95-109977-1…...
  • Page 27 High Temp. Wash Use the up and down arrows to select the desired dosing setting for the cycle Set HIP Dosing selected. (Do Not change default setting unless directed by SciCan) Prewash Press default to return to factory default. Press the back arrow to return to the Technician Menu.
  • Page 28 Press default to return to factory default. Default 0 Press the back arrow to return to the Technician Menu. P2 & P3 # extra rinses Min Value 0 – Max Value 4 Default Value P2 & P3 – 0 Hydrim L110w Service Guide 95-109977-1…...
  • Page 29 Default: Off Repeater mode will reset to OFF when Power Switch is turned OFF or set to OFF from this screen. When wash cycle is aborted by pressing the stop button, repeater mode will stay ON. Hydrim L110w Service Guide 95-109977-1…...
  • Page 30 SW Upgrade option allows technician to upgrade software by using a USB Data Logger and files stored on a USB Memory Stick. SW Upgrade USB Press the back arrow to return to the previous menu. Hydrim L110w Service Guide 95-109977-1…...
  • Page 31: Troubleshooting Using Technical Service Menu

    Drain pump is OFF rotate) Salt valve is OFF Salt switch indicates No Dosing pump is OFF Salt View IO status: This screen should be used when testing components and wires for functionality without running a cycle. Hydrim L110w Service Guide 95-109977-1…...
  • Page 32: Cycle Faults

    Defective chamber full switch Defective pressure switch Defective circulation pump CF3 Chamber Temperature Sensor Failure Detection Temperature sensor readings are out of limits (either low or high). Cause Broken temperature sensor wire, bad temperature sensor connection to PCB. Hydrim L110w Service Guide 95-109977-1…...
  • Page 33 OFF after a 15 second timeout. Cause Flow switch stuck ON. CF15 Chamber Overflow Detection The overflow switch did not turn OFF after 30 seconds of the drain pump running. Cause Defective overflow switch Clogged drain Hydrim L110w Service Guide 95-109977-1…...
  • Page 34 Side and Rear Covers To remove the side and rear covers, remove top cover first then lift up on side or rear covers and remove cover. Note: When replacing/re-attaching cover, ensure that ground wires are re-attached Hydrim L110w Service Guide 95-109977-1…...
  • Page 35: Component Replacement

    U plate from outside. After removing the mounting nuts and U plate the Water Heater is removed from inside the chamber underneath the filter screen. Remove safety switch probe before removing the heating element. Water Heater Part # 23012.01 Hydrim L110w Service Guide 95-109977-1…...
  • Page 36 Remove the screw above the touch screen and open door. Disconnect I/O Board and Logic board from Touchscreen, replace the touchscreen and re-connect the boards in reverse order. Close the door and re-insert the screw above the touchscreen. Touch Screen Part # 01-109785S Hydrim L110w Service Guide 95-109977-1…...
  • Page 37 Power Supply Board Part # 01-109782S Power Supply Board Dryer Motor and Heater Remove top cover and right side panel to access the Dryer Motor and Heater Assembly. Dryer Motor & Heater Part # 01-110218S Dryer Motor and Heater Assembly Hydrim L110w Service Guide 95-109977-1…...
  • Page 38 Inlet Valves. The dual valve is for cold water and the single Cold Water valve is for hot water. Cold Water Inlet Valve Part # 22089.04 Hot Water Inlet Valve Part # 22089.05 Hot Water Hydrim L110w Service Guide 95-109977-1…...
  • Page 39 Remove top cover and right side panel to access the Drain Pump. Drain Pump Part # 21942.00 Drain Pump Circulation Pump Remove top cover, rear and left side panels to access the Circulation Pump. Circulation Pump Part # 21953.01 Hydrim L110w Service Guide 95-109977-1…...