5 Maintenance

5.7 Storage

5.7
Storage
General recommendation
For storage, the system must first be cleaned as described in
page
solution.
Columns and media shall be stored according to their respective instructions.
Storage conditions
The following conditions shall be maintained while the system is in storage:
After storage, clean the system, calibrate all monitors, and perform a leakage test before
using the system.
48
45. After cleaning, the system must be filled with 0.01 M NaOH or 20% ethanol
Temperature: 2°C to 30°C (preferably room temperature)
Relative humidity: 0% to 95%, non-condensing (preferably low humidity).
Cleaning-In-Place, on
ÄKTAprime plus Operating Instructions 28-9597-89 AC