Pre-Installation
Serial Number Alignment
The ST75 and ST75V (Vortab) can be specified with integral or remote electronics. The flow element has a serial number etched
into the side of the HEX as shown on Figure A. In addition, the enclosure tag also shows serial and model numbers. The associated
transmitter circuit card has a serial number noted on the board as shown in Figure B. The flow sensor and transmitter circuit have been
calibrated as a matched set and should be paired together in service unless otherwise approved by a factory technician.
Flow Direction Alignment
All sensor elements have a flow arrow indicator marked on the element assembly at the reference HEX flat. These flow elements have
been calibrated in a particular direction and are designed to be used in service with the flow arrow going in the same direction as flow in
the pipe stream. The flow element has been calibrated directly in the pipe tee or tube tee with careful consideration for orientation and
insertion depth. Removal from the tee section while physically possible is not recommended unless authorized by a factory technician.
FLOW
247331
Figure A
Recommended Straight Run
To optimize flow meter system performance, FCI recommends installation with a minimum of 20 diameters upstream straight run and 10 pipe
diameters of downstream straight run. Where straight run limitations significantly reduce the available pipe diameters, FCI uses flow conditioners
to produce a transferable flow profile from the calibration installation to actual field installations. FCI's proprietary AVAL software is available
to make flow meter installation evaluations where straight run limitations are considered. See Figure C for recommended installation.
This page is subject to proprietary rights statement on last page
ST75/ST75V MASS FLOW METER
Installation and Operation Guide
C00570-1
FLOW
DIAMETER
AT LEAST 20 DIAMETERS
AFTER LAST FLOW DISTURBANCE
OPTIMUM CONDITION
Figure C - Recommended Straight Run Condition
Flow Direction
FLOW
Serial Number
10 DIAMETERS
BEFORE NEXT
FLOW DISTURBANCE
1
FC88/CLI
P3
F1
(TR5,
FCI
1.6A,
SB, UL)
ON
-
EARTH
GND
18-36VDC POWER IN
SIGNAL OUT
C00938-1
Figure B
C00568-3-1
06EN003368 Rev. G

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ST75 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for FCI ST75

  • Page 1 Serial Number Alignment The ST75 and ST75V (Vortab) can be specified with integral or remote electronics. The flow element has a serial number etched into the side of the HEX as shown on Figure A. In addition, the enclosure tag also shows serial and model numbers. The associated transmitter circuit card has a serial number noted on the board as shown in Figure B.
  • Page 2: Specifications

    DC: 18 VDC to 36 VDC (6 watts max.) Accuracy: AC: 90 VAC to 264 VAC (12 watts max.; CE Mark Approval ST75: ±2% of reading, ±0.5% of full scale for 100 VAC to 240 VAC) Optional: ±1% of reading, ±0.5% of full scale ST75V: ±1% of reading, ±0.5% of full scale...
  • Page 3: Flow Element Installation

    Process Connections The ST75 is available in pipe tee configurations with NPT threads and tubing tees with a compression fitting suitable to clamp down on concentric smooth surface tubing. The pipe tee versions are standard 150# class rated tees suitable for service up to 150 PSIG at the process temperature maximum of 250 °F (121 °C).
  • Page 4: Instrument Wiring

    Install an input power disconnect switch and fuse near the instrument to interrupt power during installation and maintenance. Operator must have power disconnected before wiring. See the “Specific Conditions of Use” and “Safety Instructions” sections in Appendix A for use of the ST75/ST75V (AC and DC versions) in Hazardous Areas Category II (Zone 1).
  • Page 5: Pulse Output Activation

    Pulse Output Activation The ST75/ST75V provides a pulse output feature. Instruments ordered with volumetric or mass flow units are factory set with totalizer and pulse output activated. The mode can be changed in the field. Source and sink output wiring is shown in Figures D and E below.
  • Page 6: Setup Interface

    After configuring the serial port (and making the PC-to-instrument connections) start a communications session with the applicable se- rial port. Enter any of the meter’s single letter commands in the program’s terminal window to execute a function. See “Table 5. ST75/ ST75V List of Single Letter Commands”...
  • Page 7: Start-Up And Commissioning

    ST75/ST75V Mass Flow Meter Fluid Components International LLC Start-up and Commissioning 1. Verify all input power and output signal wiring is correct and ready for initial power start-up. 2. Apply power to the instrument. The instrument initializes in the Normal Operation Mode with all outputs active. An instrument with the display option indicates flow with the factory-set flow units.
  • Page 8 The user cannot exit some routines unless all entries are completed or the power is cycled. The top level of the menu is shown in “Table 5. ST75/ST75V List of Single Letter Commands” on page 23. Enter the command letter mnemonic as listed in the tables below and in Appendix B to run a command.
  • Page 9: "V" Menu Output Configuration Setup

    ST75/ST75V Mass Flow Meter Fluid Components International LLC “V” Menu Output Configuration Setup Use the V menu to set up the 4-20 mA analog outputs (including NAMUR configuration) and source/sink (pulse) outputs. Note: The display comes up to the last setting saved and stays for 2 seconds. If N or [ENTER] is entered, the menu proceeds to the next item in the sequence (refer to “Table 3.
  • Page 10: Maintenance

    If no change, select N and/or [ENTER] to continue to normal operation (programming finished). Maintenance The FCI instrument requires little maintenance. There are no moving parts or mechanical parts subject to wear in the instrument. The sensor assembly, which is exposed to the process media, is composed of 316 SS and Hastelloy C.
  • Page 11: Troubleshooting

    ST75/ST75V Mass Flow Meter Fluid Components International LLC Troubleshooting Application Verification After verifying that the flow meter is functioning, review the application parameters as shown below to verify the calibration matches the process media. Equipment Needed Flow Instrument Calibration Data...
  • Page 12 The instrument uses a set of predetermined calibration parameters to process flow signals. Most of these parameters should not change. A data package shipped with the instrument (typically on a CD-ROM) contains the ST75/ST75V Delta R Data Sheet. This contains the calibration parameters stored in the flow transmitter at the factory. To verify that these parameters have not changed, complete the following: Identify the appropriate Delta R data sheets by serial number of the instrument.
  • Page 13: Hardware Verification

    Unscrew the clear cover on the fuse and pull the fuse out of the fuse holder. Check the fuse for continuity. If fuse reads open, replace with equivalent component (FCI part no.
  • Page 14 FCI Customer Service department at 1-800-854-1993 or 1-760-744-6950. Contact FCI to obtain a Return Authorization before returning the instrument. The form contains a declaration of decontamination cleaning information that the instrument must comply with before it is shipped to FCI.
  • Page 15 Turn power OFF. Mark all sensor element wires connected to the circuit board for easier reconnection to the proper terminals. Disconnect the wires. Connect the resistance decade box to the electronics as per the appropriate wiring diagram for the ST75/ST75V. Note: Interconnecting wiring (resistance decade box to electronics) must be same gauge and length to avoid any inaccura- cies in the Delta R verification caused by unequal wire lengths and/or wire gauges.
  • Page 16 Fluid Components International LLC ST75/ST75V Mass Flow Meter INTENTIONALLY LEFT BLANK 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 17: Appendix A - Approval Information

    We, Fluid Components International LLC, located at 1755 La Costa Meadows Drive, San Marcos, California 92078- 5115 USA, declare under our sole responsibility that the ST51 Flowmeter Product Family (ST51, ST75, ST75V, ST51A, ST75A, ST75AV), to which this declaration relates, are in conformity with the following standards and Directives.
  • Page 18 Fluid Components International LLC ST75/ST75V Mass Flow Meter Specific Conditions of Use per FM16ATEX0008X The ambient temperature range and applicable temperature class of the sensor probe is based on the maximum process tem- perature for the particular application as follows; T6…T1 for T of -40 °C to +65 and T...
  • Page 19 ST75/ST75V Mass Flow Meter Fluid Components International LLC Safety Instructions for the use of the ST51/75 flow meter in Hazardous Areas Approval FM16ATEX0008X/IECEX FMG 16.0009X for: Category II 2 G for Gas protection Ex db IIC T6...T1 Gb Category II 2 D for Dust protection Ex tb IIIC T85°C...T300°C Db; IP66/IP67 The ST51/75 series consist of a sensing element and associated integral or remote mounted electronics mounted in a type “d”...
  • Page 20: English - Safety Instructions

    Fluid Components International LLC ST75/ST75V Mass Flow Meter GB IRL English - Safety instructions These safety instructions are valid for the Fluid Components, ST51/75 flow meter to the EC type approval certificate no FM16ATEX0008X/IECEX FMG 16.0009X (certificate number on the type label) for use in potentially explosive atmospheres in Category II 2 GD.
  • Page 21: Consignes De Sécurité

    Consignes de sécurité Ces consignes de sécurité sont valables pour le modèle ST51/75 de la société Fluid Components (FCI) conforme au certificat d’épreuves de type FM16ATEX0008X/IECEX FMG 16.0009X (numéro du certificat sur l’étiquette signalétique) conçu pour les applications dans lesquelles un matériel de la catégorie II2GD est nécessaire.
  • Page 22 Fluid Components International LLC ST75/ST75V Mass Flow Meter Português - Normas de segurança Estas normas de segurança são válidas para os Fluid Components, ST51/75 conforme o certificado de teste de modelo N.º FM16ATEX0008X/IECEX FMG 16.0009X (número do certificado na plaqueta com os dados do equipamento) são apropriados para utilização em atmosferas explosivas categoria II 2 GD.
  • Page 23: Appendix B - List Commands

    ST75/ST75V Mass Flow Meter Fluid Components International LLC Appendix B - List Commands Table 5. ST75/ST75V List of Single Letter Commands COMMAND COMMAND COMMAND DESCRIPTION MNEMONIC FUNCTION AvgDelta_r, AvgRef Delta_r, Ref_r Tcdelta_r, Ref_r Diagnostics Kfactors Clear FlashEE, Boxcar Count, ADC to Ohms...
  • Page 24 Fluid Components International LLC ST75/ST75V Mass Flow Meter Table 6. ST75/ST75V List of CLI Commands, Cont. COMMAND COMMAND COMMAND DESCRIPTION DATA TYPE MNEMONIC FUNCTION K[1-4] K Factors Float Line Size 0 Float Line Size 1 Float Minflow Float Maxflow Float...
  • Page 25 ST75/ST75V Mass Flow Meter Fluid Components International LLC Note: When invoking a Write Function, there must be a space separating the Command characters and the data value. All Read and Write Functions are completed with a <CR>. To exit CLI, press <CR> following the last Command <CR>.
  • Page 26 Fluid Components International LLC ST75/ST75V Mass Flow Meter INTENTIONALLY LEFT BLANK 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 27: Appendix C - Drawings

    ST75/ST75V Mass Flow Meter Fluid Components International LLC Appendix C - Drawings 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 28 Fluid Components International LLC ST75/ST75V Mass Flow Meter 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 29 ST75/ST75V Mass Flow Meter Fluid Components International LLC 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 30 Fluid Components International LLC ST75/ST75V Mass Flow Meter 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 31 ST75/ST75V Mass Flow Meter Fluid Components International LLC 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 32 Fluid Components International LLC ST75/ST75V Mass Flow Meter 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 33: Customer Service/Technical Support

    Attn: Customer Service Department By Phone Contact the area FCI regional representative. If a field representative is unable to be contacted or if a situation is unable to be resolved, contact the FCI Customer Service Department toll free at 1 (800) 854-1993.
  • Page 34 Contact an FCI field representative to request field service. A field service technician is dispatched to the site from either the FCI factory or one of the FCI representative offices. After the work is complete, the technician completes a preliminary field service report at the customer site and leaves a copy with the customer.
  • Page 35: Return Authorization Request

    Return Product Information Model No: ____________________________________ Serial No(s): ____________________________________ Failure Symptoms (Detailed Description Required) : ____________________________________________________ ____________________________________________________________________________________________ What Trouble Shooting Was Done Via Phone or Field Visit by FCI: _________________________________________ ____________________________________________________________________________________________ FCI Factory Technical Service Contact: _____________________________________________________________ Sensor Element Electronics...
  • Page 36 Certificate, and providing the MSDS, shall not waive our responsibility to provide a neutralized, decontaminated, and clean product for evaluation/repair at FCI. Cleanliness of a returned item or acceptability of the MSDS shall be at the sole discretion of FCI. Any item returned which does not comply with this certification shall be returned to your location Freight Collect and at your risk.
  • Page 37: Instrument Warranty

    ST75/ST75V Mass Flow Meter Fluid Components International LLC Instrument Warranty Goods furnished by the Seller are to be within the limits and of the sizes published by the Seller and subject to the Seller’s standard tolerances for variations. All items made by the Seller are inspected before shipment, and should any...
  • Page 38 Fluid Components International LLC ST75/ST75V Mass Flow Meter Notes 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 39 ST75/ST75V Mass Flow Meter Fluid Components International LLC Notes 06EN003368 Rev. G This page is subject to proprietary rights statement on last page...
  • Page 40 1755 La Costa Meadows Drive, San Marcos, California 92078 USA - 760-744-6950 - 800-854-1993 - Fax 760-736-6250 © Copyright 2018 by Fluid Components International LLC. All rights reserved. FCI is a registered trademark of Fluid Components International LLC. Information subject to change without notice.