Troubleshooting; Identifying And Repairing Defects - Invacare Softform Premier Active 2S User Manual

Mattresses
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Invacare® Softform® Premier Active 2S

9 Troubleshooting

9.1 Identifying and repairing defects

There are audio and visual alarms present on the control unit.
Cause
Problem / Alarm
Mattress not inflating
Mattress CPR hose disconnected.
(not alternating
Has power and fuse is good, control unit does
properly).
not come on.
Major leak in air cell.
CPR hose or tube connectors kinked or split.
Not alternating, rotor failure.
No air (control unit failure).
No power.
Control unit off.
Power cord disconnected.
No power in the power source.
Power outage.
Fuse blown.
20
Solution
Connect CPR hose connector, lock it in place.
Send control unit back to Invacare for repair.
Replace leaking air cell.
Unkink or replace split CPR hose or tube
connectors.
Send control unit back to Invacare for repair.
Send control unit back to Invacare for repair.
Check power source, turn unit on.
Connect power cord and ensure the power source
is on.
Check the power source has power and turn it on.
Wait till the power source has power.
Change fuse on power inlet connector with spare
fuse or identical replacement only (consult a
trained engineer if you are unsure how to change
a fuse).
1627426-A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents