WARNING!
This mattress is recommended to be installed on
medical bed frames with cot sides or side rails.
It is preferred that the rails to be in the raised
position whenever the patient is on the bed.
Health care professionals assigned to each case
should make the final decision as to whether side
rails are warranted after assessing patient risk
of entrapment.
Controls on the footboard may be obstructed by
the control unit on a few bed frames. It may be
necessary to relocate the control unit.
– Before placing the patient on the bed, check
that air hoses and power cord are clear of
moving bed components.
– Operate all bed frame motorized functions
through their full range of motion to be
certain that there is no pulling, interference
or pinching.
– Take care when positioning hoses and cables
to eliminate the risk of tripping hazards or
strangulation.
– Ensure that the air tubing is not obstructed or
kinked when in use/positioned.
1627426-A

4.2 Installing the system

1. Place the pump unit at the end
of the bed using the built in pump
hooks. Placing the pump on the floor
will not affect the performance, but
may expose it to accidental damage.
2. Detach the Softform®Premier Active
2S twin hose from the end of the
mattress.
3. Connect it to the control unit. (An
automatic click will signify a secure
connection).
4. Connect the mains power lead to
the control unit and a suitable power
supply.
5. Switch on mains power.
6. Switch on control unit switch.
System will pressurize shown by a
yellow light and three audible beeps.
Correct pressure is shown by a green
light.
Refer to section 5.3 Unit display, page 14 for
shutdown procedure.