High Pressure

The high pressure port is used to measure pressure more
than 200 mbars. Figure 4 shows the high pressure port on
the Product.
Figure 4. High Pressure Port
For measurements to a maximum of 200 mbar,
Fluke Biomedical recommends you use the
differential pressure port. Accuracy is 100 times
higher.
 Caution
To prevent damage to the high-pressure
sensor, do not measure pressure more than
15 bar.
High
Pressure
Note
O
Measurement Cell
2
The Product has an interface for an O
See Figure 5. For more information, see the O
section in this manual.
gyo009.eps

Controls

Table 3 is a list of the front-panel controls.
Gas Flow Analyzer
measurement cell.
2
O
– Measuring Cell
2
Figure 5. O
Cell
2
Start Up
Sensor
2
gyo010.eps
7