individual settings. Table 5 is a list of screens that show in
the display.
Table 5. Settings Screens
Screen
VT305
Owner:
Company:
Next Calib:
Last Calib:
Software:
Hardware:
Description
Information
Shows device data. You can
set Owner and Company data
fields with the browser-based
configurator. See the Product
Configuration section.
Battery
Shows the current charge of
the battery.
Table 5. Settings Screens (cont.)
Screen
Ethernet
Ethernet
The Ethernet screen is used to
Default
set the Ethernet communication
parameters.
IP:
192.168.1.1
Subnet:
255.255.255..0
Set Trigger
Trigger
The Trigger events screen is
Adult
used to set when the Product
calculates volume, and
Start:
60ms
Flow:
>3.0 l/min
respiratory parameters. The
End:
60ms
Flow:
>3.0 l/min
factory defaults show adult,
pediatric, and high frequency
trigger configuration. See the
Measure Key Respiratory Data
section.
Set Gas Standard
The Product calculates the
Standard
measured flow and volume
ATP
values for the set standard. See
Amb. Temperature/Pressure
the Gas Standard after the
Specifications section.
Gas Flow Analyzer
Operation
Description
11