Display Remaining Battery Power - Elster EK280 Operating Manual

Volume conversion device
Hide thumbs Also See for EK280:
Table of Contents

Advertisement

Maintenance
For closing the administrator lock move the cursor to the "Admin" regis-
ter and to the "St.AL" value (status administrator lock) via the following
path:
Admin.
Press the ENTER button  "open" will start to flash.
Move the cursor using the arrow keys  or  to change the value to
"locked".
After you have changed the value, press the ENTER button to confirm
the inputs. The input value can be deleted by pressing the ESC button.

7.2.3 Display remaining battery power

The remaining battery power is calculated separately from
the consumed power (which is measured) and from the an-
ticipated future consumption (giving the theoretical remain-
ing battery power). Therefore, for applications which are
very power-consuming, the remaining battery power may
drop quicker than is shown on the battery power display.
The remaining battery power will automatically be recalcu-
lated after entering a new battery capacity "Bat.C" (see
above). The value cannot be edited.
Move the cursor to the "Serv." register and to the "Bat.R" value (remain-
ing battery power) via the following path:
Serv.  Batteries  Bat.R
After entering the new battery capacity, a value of at least
60 months will be displayed in standard measurement con-
ditions (see chapter 3.2.1) for "Bat.R".