To minimize or expand the software keypad

To troubleshoot a connection

Operating Guide
the new shortcut, then click Store to save it and OK to close
the Option dialog. Shortcuts must be unique. Click Default to
restore the factory shortcut values.
Settings > Option > Log
Select the Log tab to display the log entries compiled by the
software keypad. The software logs connection events,
communication errors, and similar events.
Settings > Language
Use Settings > Language to select the language for the
software keypad user interface. After a brief pause, the UI
reloads in the new language. This setting changes only the
software keypad language, not the language of the GC.
You can also turn off language selection during program
startup by deselecting Settings > Select Language Before Startup.
Click
or
in the bottom right corner of the window
to toggle keypad display.
If the software keypad cannot connect to the GC, check the
following:
• Verify GC is turned on.
• Verify LAN cabling is connected properly.
• Verify entered IP address is correct for GC. On the GC
front panel, press
This the GCs current IP address.
• Verify basic communications to the GC by using the ping
command. See the
• Verify that no one else is currently controlling the GC.
• Make sure your PC is able to communicate with the GC.
The PC IP address must be set for a similar network and
subnet. For example, if the GC IP address reads
192.168.0.26, then your PC IP address must be
192.168.0.xx, where xx is any number from 0 to 25 or 27
to 255. If the PC is set for a different LAN than the GC,
you must change the PC IP address. Refer to Windows
help for details. This operation may require administrative
privileges on the PC.