To pause a running sequence

To resume a paused sequence

To stop a running sequence

To resume a stopped sequence

Aborting a sequence

Operating Guide
Running a Method or a Sequence from the Software Keypad
The Run indicator will light and stay lit until the sequence
is completed. The sequence continues to run until all
subsequences are executed or until the sequence is
aborted.
Press [Seq Control].
1
Scroll to Pause sequence and press [Enter].
2
The sequence stops when the current sample run is
complete.
Press [Seq Control].
1
Scroll to Resume sequence and press [Enter].
2
The sequence resumes with the next sample.
Press [Seq Control].
1
Scroll to Stop sequence and press [Enter].
2
The sequence stops at the end of the currently running
subsequence unless [Seq] > Repeat sequence is On.
Press [Seq Control].
1
Scroll to Resume sequence and press [Enter].
2
The sequence restarts from the beginning of the sequence.
When a sequence is aborted, it stops immediately without
waiting for the current run to finish.
The following will cause a sequence to abort:
• The [Stop] key is pressed.
• A sampler error occurs, producing an error message.
• The GC detects a configuration mismatch during a method
load.