To Set Up EMF for Use
Advanced Operation
To use EMF:
1
After performing maintenance on an item (replacement or
calibration, for example), set a limit for its counter. See
To Set a Limit for a
2
The counter will increment as the HS performs injections
and runs sequences.
When the counter exceeds the limit, the HS will light the
3
Service Due indicator and log an entry into the Event log.
Replace or calibrate the item.
4
Reset the item's counter. See
5
Early Maintenance Feedback
Counter.
To Reset a
7
Counter.
87