To set the LAN address at the keyboard
To use a DHCP server
Advanced Operation
Press [Options]. Scroll to Communications and press [Enter].
1
Scroll to IP. Enter the numbers of the HS IP address,
2
separated by dots, and press [Enter]. A message tells you
to power cycle the instrument. Do not power cycle yet.
Press [Clear].
Scroll to GW. Enter the Gateway number and press [Enter].
3
A message tells you to power cycle the instrument. Do not
power cycle yet. Press [Clear].
Scroll to SM and press [Mode/Type]. Scroll to the
4
appropriate subnet mask from the list given and press
[Enter]. A message tells you to power cycle the instrument.
Do not power cycle yet. Press [Clear].
Scroll to Reboot instrument. Press [On/Yes] and [On/Yes] to
5
power cycle the instrument and apply the LAN setpoints.
Note that DHCP is not supported by all Agilent data
systems. Most data systems instead require a fixed address.
By using the HS MAC address, however, you can also use
DHCP to assign a host name for the HS, and use this host
name to configure the HS in the data system.
Press [Options]. Scroll to Communications and press [Enter].
1
Scroll to Enable DHCP and press [On/Yes]. When prompted,
2
turn the instrument off and then on again.
6
Options
83