5

Configuration

Configuration

66
Settings which generally apply to all methods, or which
usually do not change often, are called configuration
parameters. A typical example of a configuration parameter
is a gas type. Selections for configuration can change the
parameters available in methods, or prevent (in some cases)
a method from running. For example, the selection for
carrier gas control mode, when available, changes the
parameters available for all methods used on the HS.
Similarly, the HS changes the available configuration choices
based on installed accessories. For example, adding a carrier
gas EPC module accessory automatically adds new
configuration choices. See
Table 11 How installed hardware changes available parameters and
choices
Hardware
G4562A, 7697A Carrier Gas EPC
Module Accessory
111 vial tray
12 vial carousel
G4561A, 7697A Barcode Reader
Accessory
G4563A, 7697A Cooling Plate
Accessory
Table
11.
When installed, find:
Choices for carrier gas control.
When not installed, only GC carrier gas
control is available.
111 vial positions for a sequence.
12 vial positions for a sequence.
Method parameters for barcodes.
Barcode sequence actions.
Method parameters for cooling the vial
racks.