Setting Up for a New Method

Advanced Operation
While the HS can run sequences that include many methods,
all methods used during a single HS sequence must have the
following in common:
• Same sample loop size
• Same HS carrier configuration (including GC column
dimensions when using HS Control or GC + HS Control for the
carrier gas)
• Same gas types
All other parameters, including vial size, can vary between
samples in the sequence.
Any sample which requires a different sample loop size, gas
type, or HS carrier configuration (including GC column
dimensions as noted above) cannot be run at the same time
as samples for that other method. Install the necessary
hardware and reconfigure the HS.
3
Method Development
57