How The Hs Reduces Carryover - Agilent Technologies 7697A Advanced Operation Manual

Headspace sampler
Hide thumbs Also See for 7697A:
Table of Contents

Advertisement

2
How the 7697A Headspace Sampler Works

How the HS Reduces Carryover

26
The 7697A HS provides two special features to reduce
carryover.
• After each vial, the HS purges the sample loop and
sample probe with a high flow of vial pressurization gas,
as defined in the method. This is called the Purge flow,
and you control both the flow rate and purge time.
• Between each sequence, the HS purges the sample loop
and sample probe with a continuous, low flow of vial
pressurization gas. This is called the Standby flow. You can
control the flow rate.
Advanced Operation