To Run a Priority Sample During a Sequence

N O T E
Operation
This feature is available only with the 111 vial model, and only when
operating in standalone mode. If using an Agilent data system, edit the
running sequence instead.
You may run only 3 priority samples during a sequence. You
should avoid using sample vial positions 109–111 if using
priority samples.
To run a priority samples during a sequence:
Check the status of the sequence. Find the vial last
1
placed into the oven. (Press [Status]. See
and Sequence
Status".) The HS will use this vial's method
to process the new sample.
2
Place the new sample vial in the next priority sample
position. Start with priority sample position 1. See
Figure
15.
Priority sample 3 (111)
Priority sample 2 (110)
Priority sample 1 (109)
Figure 15 Priority sample locations
Press [Priority Sample].
3
8
Running Samples
"To View Vial
79