Filling Sample Vials
17 mm
50% of
height
typical
20-mL and 22-mL vials
Operation
In general, fill sample vials about half way. Although sample
amounts can vary depending on the analysis, do not fill vials
more than the amount shown in
correctly ensures that the sampling probe will not contact
with the matrix during sampling. If you need more sample,
use a larger vial or optimize the method to improve results.
See the
Advanced Operation Guide
75% of
height
maximum
Figure 8
Vial fill limits
Figure
for recommendations.
50% of height
maximum
10-mL vials
5
Sample Vials
8. Filling the vial
41