Headspace Sampler Status
Alert tones
Operation
When the HS is ready to begin the current sequence, the
display screen shows a Ready status, as shown below:
STATUS - Ready
Ready for start sequence
Alternatively, when a component of the HS is not ready to
begin a run, the Not Ready indicator lights, and the status
display shows Not Ready and explains why the HS is not
ready.
STATUS - Not Ready
Oven is turned off
At any time, press [Status] to view messages explaining the
current HS state, including whether the HS is ready to begin
sample preparation or what conditions are currently
preventing it from continuing.
Sequence status: Displays any status information related to the
sequence.
Setpoint status: Displays general status information, such as
readiness messages and faults states, and also lists the
setting and current value data for HS setpoints.
Vial status: During sequence execution, displays details about
the status of the selected vial. Use the arrow keys to select
the desired vial.
One beep sounds when a problem exists, but the problem
will not prevent the HS from executing the sequence. The HS
will emit one beep and display a message. The HS can start
the sequence and the warning will disappear when another
sequence starts.
Fault messages indicate hardware problems that require user
3
The Keypad
29