To Create a Method

N O T E
54
To create a basic method, you will access the method
parameters using the 4 component keys, in order: [Temps],
[Times], [Vial], and [Carrier]. For a listing of commonly- used
parameters, see
"Method Parameter
When the HS is not processing samples, press [Temps]. The
1
display shows the current oven, loop/valve, and transfer
line temperature setpoints and actual temperatures.
Scroll to Oven.
a
Use the keypad to enter the new oven setpoint, then
b
press [Enter].
Enter the desired Loop/Valve temperature, then press
c
[Enter].
Enter the desired Transfer line temperature, then press
d
[Enter].
Temperature setting is complete.
Press [Time].
2
Scroll to GC cycle time.
a
Use the keypad to enter the new GC cycle time, then
b
press [Enter]. To determine the correct value to input,
see
"To Determine the GC Cycle
Enter the desired Vial equib time, then press [Enter].
c
Enter the desired Pres equib time, then press [Enter].
d
Enter the desired Inject time, then press [Enter].
e
Time setting is complete.
3
Press [Vial].
Scroll to Fill mode.
a
Press [Mode/Type] to display the selections for the vial
b
fill mode. Scroll to the desired fill mode, then press
[Enter].
c
Enter any vial fill mode parameters, pressing [Enter]
after each.
If using Constant volume loop fill, the HS determine loop fill. Skip steps
and e.
Summary".
Time".
Operation