Repairing the Display Unit
Repairing the Display Unit
34 Service Manual Signum

Opening the Display Unit

Warning
!
Disconnect the cable from the power source.
1 You must loosen four nuts (see the picture on the left) to open the display unit.
2 Cut through the warranty stickers (seals).
3 Open the housing.
4 Now all parts are accessible.

Closing the Display Unit

AUT23103.JPG
1 Plug all connectors back into the unit.
2 Close the display unit.
3 Reconnect the unit to line voltage.
4 Check the functions of the display unit to ensure that they are working properly.
Important note:
Once the scale has been repaired and checked to ensure its proper functioning,
new warranty stickers (seals) must be applied.