AM8010 Precision Thermometer User's Guide
Rs =
100.0038
Rt = 100.0038
6. Wait for Rt and Rs to be stable ( the status sign should flash
around 10times),then press
SAVE Rs
100.0038
Menu = No
And the second line should be flashing. Press
the new Rs. Press
After the calibration, turn the AM8010 off and change back the
Mode Switch to Measurement Mode.

7.2 System Calibration

System Calibration is based on the Instrument Calibration. It is to
integrate a particular PRT sensor with AM8010 after Instrument
Calibration to form a temperature measurement system. This
system is to be calibrated at certain temperature points to calculate
sensor's coefficients based on AM8010's readings. These
coefficients are then saved in AM8010 after calibration to achieve
better measurement performance as a system.
The advantage of System Calibration is that it can achieve higher
accuracy with lower cost because the sensor doesn't need additional
calibrations before System Calibration.
Rev. 201609V3
OHM
OHM
MENU
Enter = Yes
MENU
key to exit without saving.
,the LCD should display:
key to save
24/30