5.5 Sensor setup menu (continued)
TEST/MAINT
SET DEFAULTS
TEST/MAINT
PROBE INFO
CAL DATA
Counter for customer service settings: 0–1000 days (180 days are recommended)
Check the service contract and enter the defined value (number of days).
0 = Service deactivated
The instrument resets the settings to the default configuration.
PARAMETER: eco: NO
-N; plus, clear: NO
x
MEAS UNIT: mg/l
MEAS INTERVAL: 5 min
RESPONSE TIME: eco, plus: 3 Counts; clear: 1 Count
WIPER MODE: eco: SINGLE ; plus, clear: B-A-B, B-A-B
Select NITRATAXplus/eco/clear, LOCATION, SERIAL NUMBER, RANGE, PATH LENGTH,
WIPER P/N, MODEL NUMBER,SOFTWARE VERS, DRIVER VERS., PRODUCTION DATE
Name of connected sensor: NITRATAX plus/eco/clear
LOCATION
SERIAL NUMBER: serial number of connected sensor
RANGE: Measuring range corresponding to the measuring path
PATH LENGTH: Width of the measuring path
WIPER P/N: Item number
MODEL NUMBER: Item number
SOFTWARE VERS: Sensor software
DRIVER VERS.: STRUCTURE, FIRMWARE, CONTENT
PRODUCTION DATE: Production date
Overview of OFFSET, FACTOR, DATE, DEXT 100%, DEXT 50%, DEXT 25%, CAL, R , M, IR
and IM
OFFSET: Adjustable on the CALIBRATION menu
FACTOR: Adjustable on the CALIBRATION menu
DATE: Date of the last change of OFFSET and/or FACTOR
Internal calibration data:
DEXT 100%
DEXT 50%
DEXT 25%
CAL.: Internal calibration data
R: Internal calibration data
M: Internal calibration data
IR: Internal calibration data
IM: Internal calibration data
3
Operation
19