3.1

Respirator Fit Test

The user must perform a respirator fit test (Quantitative Test or Qualitative Test) and follow all warn-
ings and limitations specified. Failure to do so can result in serious personal injury or death.
A qualitative or quantitative respirator fit test must be carried out routinely for each wearer of this respi-
rator to determine or confirm the amount of protection that the respirator provides.
Quantitative Test
If a quantitative fit test is used, a fit factor of at least 500, based on ambient aerosol fit test methods
or equivalent, is required before any type of respirator is assigned to an individual.
Qualitative Test
If a qualitative fit test is used, only validated protocols are acceptable. The individual must pass a test
designed to assess a fit factor of at least 500.
NOTE:
When this facepiece is used for CBRN APR, a fit factor of at least 2000 is required for quan-
titative or qualitative test. Refer to User Instructions (P/N 10158743) for more information regarding
CBRN APR use.
Use Quik Chek Kit (P/N 805078) to perform fit testing and following instructions provided with the kit.
The fit test kit includes an adapter and instructions. Canister sold separately. Be sure the probe does
not contact the face during fit testing. Failure to do so can result in false protection factor readings.
WARNING
G1
Size Selection
US
20