Additional Parameters - Siemens sitrans LR 400 Instruction Manual

Hide thumbs Also See for sitrans LR 400:
Table of Contents

Advertisement

Level damping (F = 1 s)
Damping of level in s
Value
Set the damping of the level value in seconds. It acts on the analog output, the limit value
monitor and the local display. For damping of the sensor signal, set Parameter 4.2.3.
Application Type (F - Liquid [process])
Use of the vessel
Value
Select Silo1 (solids) for tall, narrow silos. Select Silo 2 (solids) for large diameter silos,
typically used for cement. In most cases you set one of the pre-specified applications
here. The user vessels may adopt configurations which deviate from the factory settings.
These are designed for special applications loaded at the factory or by service. The
following parameters cannot be accessed when you set a user vessel: Parameter 4.2.2.2,
Parameter 4.2.2.5, Parameter 4.2.2.6, Parameter 4.2.3.1 and Parameter 4.2.3.5.

Additional Parameters

2. Display
2.1: Multi-display (F = level in m
Display of two measured values. Values are determined in Parameter 4.5. 1 .1 (Line 1 Local)
and in Parameter 4.5.1.4 (Line 2 Local).
2.2: Level (F = m)
Current level of measured medium (set unit using Length Unit in Auto-Setup)
Page 32
numerical value
Liquid (store)
Liquid (process)
Silo1 (solids)
Silo2 (solids)
User tank1
User tank2
Signal to noise ratio in db)
SITRANS LR 400 – INSTRUCTION MANUAL