UEi EAGLE C125 Quick Start Manual page 4

Sensor calibration & replacement

Advertisement

G
C
AS
ALIBRATION
It is recommended to allow the sensors to stabilize a minimum of four hours before calibration.
1. Turn instrument on in fresh air
and allow to zero.
5. Select "Cal_CO".
9. Allow to sample gas three
minutes.
a. Reading does not indicate
ppm value.
b. Slight fluctuation of display
is normal.
COPYRIGHT © 2007 UEi. EAGLE Combustion Analyzers™ is a trademark of UEi.
2. Select "Menu".
6. Enter gas concentration.
10. Press "Enter" at the end of
3 minutes.
Note reading is not ppm,
and may still fluxuate slightly.
3. Scroll to "Service".
7. Connect Gas.
11. Scroll to "Exit" and then
press "Enter".
Visit us at www.TestEquipmentDepot.com
4. Enter user code "1234".
8. Adjust regulator to match flow
of instrument.
a. Floating ball should be in
the middle of the second
space from the bottom.
b. Note the pitch of the pump.
It should match the pitch
when drawing in ambient
air.