Specifications

Operating Specifications

The specifications for the standard pressure cell concentric cylinders are:
Stator inner radius:
Rotor outer radius:
Cylinder immersed height:
Gap:
Backoff distance:
Geometry inertia:
Sample volume:
Temperature range:
Maximum applied pressure:
Maximum pressure (self-pressurizing):
Torque range
Maximum angular velocity:
Seal construction:

Safety Specifications

Over pressure rupture disk:
Hydraulically tested to:

Operational Limits

CAUTION: To prevent sample entering the upper part of the cell and contaminating the
bearings, the cell should not be used above the limits given below. Exceeding these limits
may also cause mechanical damage to the cell.
Maximum Angular Velocity:
Maximum Sample Viscosity:
Maximum Frequency:
NOTE: The quality of data obtained using the pressure cell cannot be expected to match that
obtained when conventional measuring systems are used with the rheometer. Some of the
normal calibration routines are not relevant to, or cannot be used with the Pressure Cell.
Alternative calibration routines are described in this chapter.
86
172 bar (2500 psi)
414 bar (6000 psi)
50 rad/s
The geometry should not be forced into the sample. Light hand pressure
should be all that is required.
50 Hz (314 rad/s)
14.00 mm
13.00 mm
44.00 mm
3500 µm (recommended)
3500 µm
92.00 µN.m.s2 (approximate)
9.5 ± 0.5 ml
-10 to 150°C
138 bar (2000 psi)
5 bar (72.5 psi)
about 100 µN.m to 0.2 N.m
50 rad/s
DuPont Kalrez®