Replacing The Device; Return Procedure - Siemens 7ME4613 Installatio And User's Operating Instructions

Coriolis flowmeters with hart
Hide thumbs Also See for 7ME4613:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
WARNING
Hot, toxic or corrosive process media
Danger of injury during maintenance work.
When working on the process connection, hot, toxic or corrosive process media could be released.
● As long as the device is under pressure, do not loosen process connections and do not remove any parts that are
pressurized.
● Before opening or removing the device ensure that process media cannot be released.
WARNING
Improper connection after maintenance
Danger of explosion in areas subject to explosion hazard.
● Connect the device correctly after maintenance.
● Close the device after maintenance work.
6.4

Replacing the device

CAUTION
Corrosive substances
Danger of chemical burns when replacing the sensor.
The sensor in the device contains corrosive substances that result in burns on unprotected skin.
● Make sure that the sensor enclosure is not damaged when replacing the sensor.
● If contact with the corrosive substances occurs, rinse the affected skin immediately with large amount of water to dilute
substance.
6.5

Return procedure

Enclose the bill of lading, return document and decontamination certificate in a clear plastic pouch and attach it firmly to the
outside of the packaging.
Required forms
● Delivery note
● Return goods delivery note (http://www.siemens.com/processinstrumentation/returngoodsnote)
with the following information:
– Product (item description)
– Number of returned devices/replacement parts
– Reason for returning the item(s)
● Decontamination declaration (http://www.siemens.com/sc/declarationofdecontamination)
With this declaration you warrant "that the device/replacement part has been carefully cleaned and is free of residues.
The device/replacement part does not pose a hazard for humans and the environment."
If the returned device/replacement part has come into contact with poisonous, corrosive, flammable or water-
contaminating substances, you must thoroughly clean and decontaminate the device/replacement part before returning it
in order to ensure that all hollow areas are free from hazardous substances. Check the item after it has been cleaned.
Any devices/replacement parts returned without a decontamination declaration will be cleaned at your expense before
further processing.
The forms can be found on the Internet as well as in the documentation which comes with the device.
SITRANS FC430 with HART (7ME4613, 7ME4603, 7ME4623, 7ME4713)
A5E03736902-002, 12/2013
15