Error Messege
Finder port is open.
The measured object does
not exhibit fluorescence.
The fluorescent calibration
coefficients could not
be determined.
No response from instrument Communication with
Detected instrument
disagrees with connected
one.\nConnect again.
File is different to
current status.
UV adjustment not
acceptable.
Problem / possible cause
The viewfinder cover stays
open during measurement.
UV adjustment is not allowed
due to the UV specimen not
containing fluorescence.
The fluorescent
calibration coefficient
cannot be determined.
the instrument failed.
• The power of the
instrument is off.
• COM port setting
is incorrect.
• Communication
settings are incorrect.
The instrument is incorrect.
The measurement
conditions in the fluorescent
calibration coefficient file
(.krd) designated to be read
are different from the actual
measurement conditions.
UV adjustment with a
Ganz&Griesser option has
not yet included all the
required measurements.
26
Action
Close the viewfinder cover
before measurement.
Perform UV adjustment
using a specimen that
contains fluorescence.
Change Standard value
or tolerance settings,
or use a different
Fluorescent Standard;
then retry "UV adjustment".
Turn on the power of
the instrument.
Check and correct
the COM port and
communication settings.
Connect CM-3600A.
Adapt the actual
measurement conditions
to the measurement
conditions in the .krd
file or select a file that
corresponds with the actual
measurement conditions.
Measure the fluorescent
sample four times when
selecting "Ganz&Griesser 4"
or five times when selecting
"Ganz&Griesser 5".
Refer to page
22
15
SpectraMagic™
NX (optional)
instruction
manual.
SpectraMagic™
NX (optional)
instruction
manual.