Wellion Galileo GLu/KET Owner's Manual page 23

Blood glucose and ketone monitoring
Table of Contents

Advertisement

after use.
Step 5: Hold the meter and touch the control solution to the
edge of the sampling end of the test strip. The control solution
will be automatically drawn into the reaction area of the test
strip.
Step 6: The screen will begin to count down. After
G l u
5 seconds, the control solution testing result will appear on
when results are shown on the screen.
Step 7: Compare the reading on the screen to the control range printed on the test
strip bottle. If the reading does not fall within the control range printed on the test
strip bottle, see the chapter "Control Solution Trouble Shooting".
Step 8: Remove the used test strip either by pushing the ejection button (if available)
or by pulling it out with your hand. The meter will turn off automatically. The meter will
also time-out after 2 minutes of inactivity.
Perform other levels of control solution testing by repeating the above steps.
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents