Page 2
Dear Wellion CALLA Classic Owner, Thank you for choosing the Wellion CALLA Classic Blood Glucose Monitoring System to help you monitoring your blood sugar levels. We designed the Wellion CALLA Classic Blood Glucose Monitoring System to be accurate and easy to use.
Page 3
Testing is done outside the body (in vitro diagnostic use). The test results are plasma-calibrated for easy comparison to lab results. The Wellion CALLA Classic Blood Glucose Monitoring System is a portable battery operated meter intended for use as an aid by persons with diabetes and by health care professionals to monitor glucose concentration in whole blood.
TABLE OF CONTENTS YOUR NEW BLOOD GLUCOSE KIT The Wellion CALLA Classic Meter The Wellion CALLA Blood Glucose Test Strip Meter Display Screen INSERTING (OR CHANGING) BATTERIES CONTROL SOLUTION TESTING Control Solution Trouble Shooting TESTING YOUR BLOOD GLUCOSE Preparing your Lancing device...
Page 6
VIEWING YOUR METER’S MEMORY CARING FOR THE METER Cleaning the Meter Storage and Precautions SOLVING PROBLEMS MANUFACTURER´S WARRANTY SPECIFICATIONS SYMBOLS...
YOUR NEW BLOOD GLUCOSE KIT The Wellion CALLA Classic Blood Glucose Monitoring System kit includes the following items: Wellion CALLA Classic blood glucose meter 10 Wellion CALLA test strips Wellion Lancing device 10 Wellion Lancets Compact Pouch Easy Start Guide...
The Wellion CALLA Classic Meter Test Strip Port Insert test strip to turn on the meter. Meter Display Shows your blood glucose test results and other messages. Light Button on the light. meter.
Page 9
The Wellion CALLA Classic Meter Battery Compartment...
The Wellion CALLA Test Strips are glucose specific, biosensor-based test strips that can test glucose in capillary whole blood in as quickly as 6 seconds and require very little blood sample. The No Code feature eliminates error in testing caused by incorrect coding or calibration procedure.
Page 11
IMPORTANT: the Wellion CALLA Classic Blood Glucose Meter. Other brands of test strips will not work with the Meter. and light, it is important to close the vial cap of the test strip bottle tightly after each use. DO NOT leave any test strips outside the bottle...
INSERTING (OR CHANGING) BATTERIES Batteries need to be inserted before using your Wellion CALLA Classic Me- ter for the first time or when the battery icon appears on the meter display. Materials you will need: Your Wellion CALLA Classic Meter...
CONTROL SOLUTION TESTING The purpose of the control solution testing is to make sure the Wellion CALLA Classic Blood Glucose Monitoring System is working properly. You should perform control solution testing when:...
Page 16
CALLA Test Strips. Other brands of control solution will produce inaccurate results. discard any unused control solution three months after opening. Materials you will need: Your Wellion CALLA Classic Meter Wellion CALLA Control Solution A new Wellion CALLA Blood Glucose Test Strip...
Page 17
Step 1: Insert a Wellion CALLA Blood Glucose Test Strip into the meter with the arrow mark facing up. Make sure the test strip contact points are inserted all the way into the meter. Close the test strip bottle immediately after you take out a strip.
Page 18
open it. 6 months after first opened date, discard the bottle and any remaining Test Strip. 10 to 15 minutes for it to reach the new temperature before use. Your Meter will not begin testing if it’s outside this temperature range. Step 2: blood drop icon.
Page 19
Step 3: Squeeze a drop of Wellion CALLA Control Solution onto a clean, dry, nonabsorbent surface. Do not apply control solution to the test strip directly from the bottle. Replace the bottle cap on the control solution bottle immediately after use.
Page 20
Step 6: Compare the reading on the screen to the corresponding control range printed on the test strip bottle or its package. If the reading does not fall within the ranges printed on the test strip bottle or its package, see Control Solution Trouble Shooting on page 21.
NOTE: Control Solution Trouble Shooting If your control solution testing is out of range (too high or too low), it maybe caused by the following: Possible Causes What you can do ... Solution being used Control Solution. room temperature...
Page 22
Possible Causes What you can do ... control solution date on all control solution and test strips. Repeat testing using a new test strip. If the result is still out of range, use a new bottle of control solution and retest. dealer.
After every turning on of the meter it will perform a system check automa- tically. Materials you will need: Your Wellion CALLA Classic Meter A new Wellion CALLA Blood Glucose Test Strip Lancing device with a clean, unused lancet Before you begin, make sure: and dry well.
Preparing your Lancing device Step 1: Unscrew the adjustable cap of the lancing device and insert the lancet by push down firmly until it is fully seated.
Page 25
Step 2: clockwise the adjustable cover back onto the lan- cing device. NOTE: Step 3: Adjust the puncture setting on the adjustable cap for the puncture depth level (1 is the lightest and 5 is the deepest).
Kappe nochmals so weit nach hinten, bis der Auslöseknopf rot wird. Performing a Blood Glucose Test Step 1: Take out a Wellion CALLA Test Strip from the test strip bottle and close the bottle immediately. Insert the test strip to turn on the meter.
Page 27
IMPORTANT: a dry, cool place. any remaining test strip after 6 months from date of opening.
Page 28
fingertip and press the release button to create a puncture. TIP:...
Page 29
Step 3: Gently bring the test strip and touch the drop of blood at a slight angle. The test strip acts like a straw to pull the blood in. Keep the test strip in the blood drop until the meter beeps to indicate the test strip has enough blood to test.
Page 30
CAUTION: 600 mg/dL or below 20 mg/dL. Test again using fingertip testing, DO NOT test on forearm. If you still receive the same result, call your physician or healthcare professional immediately. Step 5: Remove and discard the used test strip into proper wastebasket. The meter Step 6: Remove the used lancet from your lancing device.
Page 31
with the thumb and simultaneously pull out the sliding barrel to dispose the used lancet. Discard the lancet into proper wastebasket. IMPORTANT: Discard used test strips and lancets according to your local disposal regulations where applicable.
UNDERSTANDING YOUR BLOOD GLUCOSE TEST RESULT Blood glucose value will vary depending on food intake, medication, health, , and It is important to consult with your physician or healthcare professional to determine an appropriate target range for you. Reference: 1. Stedman,Thomas Lathrop. Stedman’s Medical Dictionary, 27th Edition, 1999, pg. 2082...
VIEWING YOUR METER’S MEMORY The CALLA Classic Meter automatically stores the last test result. Materials you will need: Your Wellion CALLA Classic Meter Step 1: Turn on the meter by press and hold the light button until the screen will light up showing the last test result.
CARING FOR THE METER keep your meter working properly. Cleaning the Meter with a mild detergent. meter or hold it under running water. Storage and Precautions could damage the electronics.
Page 36
as bathroom or kitchen. sure it’s closed tightly. are technical problems or questions, call our authorized dealer for assistance.
SOLVING PROBLEMS This section details the significant display screen messages and error codes you will encounter when using your Wellion CALLA Classic meter and test strips. What you see What it means What should you do Damaged, wet, or ter.
Page 38
moved too early strip. numbers found on the packaging. tions: Big tempe- 10 to 15 minutes for it to reach the new temperature. the meter again. numbers found on the packaging.
Page 39
page 13, Inserting (or changing) the Battery. performed outside of operating tem- 10 to 15 minutes for it to reach the new temperature. higher than peat the test using a new test strip. physician or healthcare professional immediately.
Page 40
lower than peat the test using a new test strip. If the result is still LO, contact your physician or healthcare professio- nal immediately.
MANUFACTURER´S WARRANTY The producer warrants Wellion CALLA Classic blood glucose meter against defects in materials and workmanship for a period of two years from the date of purchase. The warranty is lost if the instrument is misused, poorly main- tained or is tampered with.
SPECIFICATIONS Test Strips: Wellion CALLA Blood Glucose Test Strips Test Range: 20-600 mg/dl Calibration: Plasma Test Time: 6 Seconds Sample Size: 0,65 μl Display Type: Large LCD screen Memory: The last test result Weight: 68g (incl. batteries) Battery: 2 Alkaline AAA Batteries...
Page 43
Alternate Site Testing (AST) on palm or forearm possible.
SYMBOLS (LOT) Batch Code C ˚ Temperature Limitations C ˚ F ˚ F ˚ Consult Instructions for Use Caution, consult accompanying documents Catalogue number Keep away from sunlight/direct light...
Page 45
Do not reuse Do not use if package is damaged Manufacturer FOR SELF- In vitro diagnostic, medical device for self-testing TESTING...
Page 48
Gewerbepark 10 7221 Marz In Vitro diagnostisches, AUSTRIA medizintechnisches Gerät www.medtrust.at...
Need help?
Do you have a question about the CALLA classic and is the answer not in the manual?