Chauvin Arnoux C.A 6501 User Manual
Chauvin Arnoux C.A 6501 User Manual

Chauvin Arnoux C.A 6501 User Manual

Megohmmeter
Hide thumbs Also See for C.A 6501:

Advertisement

Quick Links

MEGOHMMETRE
MEGOHMMETER
MEGOHMMETER
MEGAOHMMETRI
MEGAOHMETRO
FRANÇAIS
E N G L I S H
D E U T C H
I T A L I A N O
ESPAÑOL
Notice de fonctionnement
User's manual
Bedienungsanleitung
Manuale d'uso
Manual de instrucciones
C.A 6501
El modelo Chauvin Arnoux
C.A 6501 es el equivalente al
modelo AEMC 6501

Advertisement

Table of Contents
loading

Summary of Contents for Chauvin Arnoux C.A 6501

  • Page 1 MEGOHMMETRE C.A 6501 MEGOHMMETER MEGOHMMETER El modelo Chauvin Arnoux MEGAOHMMETRI C.A 6501 es el equivalente al MEGAOHMETRO modelo AEMC 6501 FRANÇAIS Notice de fonctionnement E N G L I S H User’s manual D E U T C H Bedienungsanleitung I T A L I A N O Manuale d’uso...
  • Page 2 English You have just purchased a CA 6501 megohmmeter; thank you for your confidence. For best results from your device: • read this user manual attentively, • observe the precautions for its use. MEANINGS OF THE SYMBOLS USED AC – Alternating current Instrument fully protected by double insulation.
  • Page 3: Table Of Contents

    1mADC: ..........19 Ω measurement at 5mADC: ..........20 3.2.3 3.2.4 AC voltage measurement ..........20 GENERAL CHARACTERISTICS: ...........21 WARRANTY ..................23 MAINTENANCE ................23 REPLACING THE FUSE: ............23 STORAGE: ................24 CLEANING: ................24 METROLOGICAL CHECK............24 REPAIR: ...................24 TO ORDER ..................25 C.A 6501...
  • Page 4: Presentation

    Red indicator lit: multiply the reading Retractable crank by 10 to obtain the result (on MΩ, kΩ, Ω scales) Screw for adjustment Range switch; of the mechanical zero set back to of the galvanometer MΩ/V∼ after using C.A 6501...
  • Page 5: Use - Characteristics

    A red indicator (Fig. 2) on the left side of the instrument indicates the automatic change of range. When it is lit, multiply the reading by 10 to obtain the result. C.A 6501...
  • Page 6: Operation

    1mADC, from 0 to 500kΩ measured (650V max. no-load). There is a change of reading coefficient for rising values: 50kΩ (changes to X10 reading), and a change of reading coefficient for falling values: 45kΩ (changes to X1 reading). C.A 6501...
  • Page 7: Ω Measurement At 5Madc

    9Ω (changes to X1 reading). 3.2.4 AC voltage measurement The AC voltage measurement is from 0 to 600V in a single range (input impedance: 100kΩ), switch set to V. C.A 6501...
  • Page 8: General Characteristics

    ± 2.5% of full scale kΩ Ω ± 2.5% of full scale ± 3% of full scale from 45 to 450Hz Volt~ Measurement voltage: 500V MΩ Short-circuit current ≤ 5mA MΩ Current regulated to 1mA kΩ Ω Current regulated to 5mA C.A 6501...
  • Page 9 ± 5% MΩ Overloads - protection: Ω Ω Ω Ω range: 600VAC/DC max.: protection by fast-blow fuse F 0.2A, 600V, HBC (6.3x32mm) from 100V, below 100V overload limited to 10 seconds. kΩ Ω Ω Ω range: 600VAC/DC max. C.A 6501...
  • Page 10: Warranty

    Using a flat blade screwdriver, press and turn a quarter turn anti-clockwise to release the fuse. Replace only with a fuse having identical characteristics. Fuse: F 0.2A/600V/HBC P/N: P01297095 C.A 6501...
  • Page 11: Storage

    Repair, under warranty or not Send your instrument to one of the CHAUVIN ARNOUX-approved MANUMESURE Regional Technical Centres Information and coordinates on request: Tel.: +33 (0)2 31 64 51 43 Fax.: +33 (0)2 31 64 51 09...