Ac current probe

Advertisement

2.56"
(65 mm)
.83"
(21 mm)
MN254

ELECTRICAL SPECIFICATIONS

Nominal Range: 20A, 200A
Measurement Range: 20A: 0.1 to 24A
Transformation Ratio: Voltage output
20A: 100mV/A
Output Signal:
200A: 10mV/A
Accuracy and Phase Shift*:
20 A Range
Accuracy: 1% ± 50mV
Phase Shift: Not specified
Overload: 24 A for 10 min ON, 30 min OFF
2.00"
(51 mm)
MN255
200A: 0.1 to 240A
.73"
(18.5 mm)
Tactical
Barrier
(keep hands below
this barrier)
.1.18"
200 A Range:
Accuracy: 0.5 to 10A: 3.0% Reading ± 5mV
10 to 40A: 2.5% Reading ± 5mV
40 to 100A: 2.0% Reading ± 5mV
100 to 240A: 1.0% Reading ± 5mV
Phase Shift: 0.5 to 10 A: Not Specified
(*Reference conditions: 18°C to 28°C, 20 to 75% RH,
external magnetic field <40 A/m, 48 to 65Hz sine wave,
distortion factor less than 1%, no DC component, no
external current carrying conductor, test sample centered.)
Load impedance > 1M.
Black
(30 mm)
10 to 40A: ≤ 5°
40 to 100A: ≤ 3°
100 to 240A: ≤ 2.5°
Red