ELECTRICAL SPECIFICATIONS

Nominal Range: 200A
Measurement Range: 0.5 to 240A
Transformation Ratio: 1000:1
Output Signal: 1mA/A on 1
Accuracy and Phase Shift*:
Accuracy:
0.5 to 10A: 3.0% Reading ± 0.5A
10 to 40A: 2.5% Reading ± 0.5A
40 to 100A: 2.0% Reading ± 0.5A
100 to 240A: 1.0% Reading ± 0.5A
2.00"
(51 mm)
AC
.73"
(18.5 mm)
Tactical
Barrier
(keep hands below
this barrier)
.1.18"
Phase Shift:
0.5 to 10A: Not Specified
10 to 40A:  5°
40 to 100A:  3°
100 to 240A:  2.5°
*Reference conditions: 18°C to 28°C, 20 to 75% RH,
external magnetic field <40 A/m, 48 to 65 Hz sine
wave, distortion factor less than 1%, no DC
component, no external current carrying conductor,
test sample centered. Load impedance 1
Overload: 240A for 10 min ON, 30 min OFF
Frequency Range: 40 to 10kHz