Technical Data

3 Technical Data

3.1 Parameters

Temperature range
Spectral response
Thermal Parameters
Accuracy
(Digital and Voltage)
1
Accuracy
(TC)
3
Repeatability (Digital and Voltage)
Repeatability (TC)
Response time (95%)
Temperature resolution
Emissivity
Transmissivity
Electrical Parameters
Power
Analog Output
Digital Output
1
for ambient temperature 23°C (73°F) ± 5 K , e = 0.95 and calibration geometry
2
± 3.5°C for T
< 0°C (32°F)
meas
3
for ambient temperature 23°C (73°F) ± 5 K , e = 0.95 and calibration geometry
± 5.5°C for T
< 0°C (32°F)
4
meas
11
-20 to 500°C (-4 to 932°F)
8 to 14 μm
± 1.5% of reading or ± 2°C
whichever is greater
± 1.5% of reading ±2°C or ± 4°C
whichever is greater
± 0.5% of reading or ± 1°C
whichever is greater
± 0.5% of reading ± 1°C or ± 2°C
whichever is greater
150 ms
0.1°C (0.2°F)
0.100 to 1.100 (software controlled)
0.100 to 1.000 (software controlled)
24 VDC ± 20% @ 20 mA
0 to 5 V or TCJ or TCK output
Two-way RS232 digital output
9600 baud, 8 data bits, 1 stop bit,
no parity, no flow control
Rev. B 06/2011